登录    注册    忘记密码

期刊文章详细信息

FY-3E微波湿度计发射前定标的非线性系数在星上定标中的适用性分析  ( EI收录)  

Analysis on applicability of nonlinearity coefficients derived from prelaunch calibration tests to onboard calibration of Microwave Humidity Sounder(MWHS-Ⅲ)on FY-3E satellite

  

文献类型:期刊文章

作  者:王振占[1] 肖雨伟[1,2] 张升伟[1] 何杰颖[1] 谷松岩[3,4,5]

WANG Zhenzhan;XIAO Yuwei;ZHANG Shengwei;HE Jieying;GU Songyan(Nation Space Science Center,Chinese Academy of Sciences,Key Laboratory of Microwave Remote Sensing,Chinses Academy of Sciences,Beijing 100190,China;University of Chinese Academy of Sciences,Beijing 100049,China;National Satellite Meteorological Center(National Center for Space Weather),China Meteorological Administration,Beijing 100081,China;Key Laboratory of Radiometric Calibration and Validation for Environmental Satellites,China Meteorological Administration,Beijing 100081,China;Innovation Center for FengYun Meteorological Satellite(FYSIC),Beijing 100081,China)

机构地区:[1]中国科学院国家空间科学中心中国科学院微波遥感技术重点实验室,北京100190 [2]中国科学院大学,北京100049 [3]国家卫星气象中心(国家空间天气监测预警中心)中国气象局,北京100081 [4]中国遥感卫星辐射测量和定标重点开放实验室中国气象局,北京100081 [5]许健民气象卫星创新中心,北京100081

出  处:《遥感学报》

基  金:国家自然科学基金面上项目(编号:4177012050);国家重点研发计划(编号:2018YFB0504900,2018YFB0504902)。

年  份:2023

卷  号:27

期  号:10

起止页码:2327-2336

语  种:中文

收录情况:AJ、BDHX、BDHX2020、CSCD、CSCD2023_2024、EI、IC、JST、RCCSE、SCOPUS、ZGKJHX、核心刊

摘  要:星载微波湿度计发射前定标和在轨定标的最大不同是冷源的使用:在轨定标的冷源是宇宙冷空背景温度,通常为2.7 K,而发射前定标的冷源通常控制在80—100 K。这样通过发射前定标得到的非线性系数是否可以用于在轨定标一直是一个非常关心和值得研究的问题。本文通过在FY-3E微波湿度计发射前热真空定标中除了传统需要的冷热、变温定标源之外,另外设计了一个独立的位于180—200 K附近的定标源。利用该验证源,本文提出了三点定标公式,基于冷空不变性并通过外推到冷空亮温的模拟观测计算非线性系数。随后通过比较传统和新型非线性系数推导方法的结果,并将两组系数带入同一组数据中分析两组发射前定标系数用于在轨定标的适用性。结论表明,通过对传统方法和新方法的计算结果比较,发现将两组非线性方法应用于同一数据集的结果存在差异,说明传统方法得到的发射前系数存在误差。并且该方法简化了发射前标定过程,提高了非线性系数用于星上标定的精度。

关 键 词:微波湿度计定标  FY-3微波湿度计  发射前定标  在轨定标  非线性系数  

分 类 号:TP751] P2]

参考文献:

正在载入数据...

二级参考文献:

正在载入数据...

耦合文献:

正在载入数据...

引证文献:

正在载入数据...

二级引证文献:

正在载入数据...

同被引文献:

正在载入数据...

版权所有©重庆科技学院 重庆维普资讯有限公司 渝B2-20050021-7
 渝公网安备 50019002500408号 违法和不良信息举报中心